T2710 is the digital portable calibrator for micro flow of ForTest® T-Series.
Enclosed in a container of anodized aluminium with shatterproof plexiglass front, we find a completely new circuit board and a mass flow meter sensor, for direct measurement of leaks.
Coupled to the micrometer nozzle supplied, they are designed to simulate the desired leakage, facilitating the procedure for programming and verification of leak test instruments.
T2710 Standard version has got a lithium ion battery rechargeable that gives it an incredible autonomy and a USB port for PC connection in order to download and manage the taken measures.
T2710 uses a mass flow sensor and can detect flow rates up to 200 cc/min with precision and reliability. This portable instrument, of compact dimensions and simple use is essential to check the calibration and the repeatability of measure instruments and of nozzle sample.
Applications Fields
Model | 0-30cc/min | 0-200cc/min | ||||
Dp / flow accuracy | 0,5% RDG+0,07%FS | 0,5% RDG+0,1%FS | ||||
Dp / flow measurement range | 0-30cc/min | 0-200cc/min | ||||
Dp / flow resolution | 0,01cc/min | 0,1cc/min | ||||
Filling pressure accuracy | - | - | ||||
Filling pressure measurement range | - | - | ||||
Filling pressure resolution | - | - | ||||
Leak measurement accuracy in cc/’ | - | - | ||||
Leak measurement range in cc/’ | - | - | ||||
Leak measurement resolution in cc/’ | - | - | ||||
Volume measurement accuracy in cc/’ | - | - | ||||
Volume measurement range in cc/’ | - | - | ||||
Volume measurement resolution in cc/’ | - | - | ||||
Unit of measurement | cc/min, cc/h, °C | |||||
Dimensions | 180X80X35 mm | |||||
Weight | 0,7 kg | |||||
Display dimensions | 128x64 pixel | |||||
Working temperature range | 5-40 °C | |||||
Port RS232 | - | |||||
Port RS485 | - | |||||
“Master” USB port | - | |||||
“Slave” USB port | Yes | |||||
Ethernet port | - | |||||
I/O signals | - | |||||
Auxiliary I/O signals (optional) | - | |||||
Number of programs | - | |||||
Test archive memory | - | |||||
Lock with password | - | |||||
Program name | - | |||||
Note | The resolution to be considered for accuracy is always 1 Pa. Reference Normatives: EN 61010-1, EN61326-1 / EN61326/A1, EN61000-3-2 / EN61000-3-2/A14, EN61000-3-3 / EN61000-3-3/A1, EN61000-4-2 / EN61000-4-2/A1, EN61000-4-3 / EN61000-4-3/A1 / EN61000-4-3/A2, EN61000-4-4, EN61000-4-5, EN61000-4-11, EN1779
|